5 Hygienic reprocessing

5
Hygienic reprocessing

5.1 General instructions

5.2 Intervals

74
EN
ACCUVAC Pro/ACCUVAC Lite
This product may contain disposables. Disposables are
intended to be used only once. So use these items only once
and do not reprocess them Reprocessing disposables may
impair the functionality and safety of the product and lead to
unforeseeable reactions as a result of ageing, embrittlement,
wear, thermal load, the effects of chemical processes, etc.
Wear suitable gloves for disinfection work (e.g., household or
disposable gloves).
Please refer to the instructions for use supplied with the
disinfectant used.
Also observe the respective instructions for use for the
accessories and the other parts.
The service life of reusable components is a maximum of
50 reprocessing cycles.
Part
Device, accessories and
other parts
Hygiene filter for
disposable canister
system
Intervals
After every patient
Immediately in the case of discoloration,
soiling or oversuction
or
If the vacuum is > -0.2 bar while at maximum
suction power with an open suction hose
or
Every 30 days