9 Hygienic reprocessing

9
Hygienic reprocessing

9.1 General instructions

144
EN
MEDUCORE Standard
This product may contain disposable items. Disposable
items are intended to be used only once. So use these
items only once and do not reprocess them. Reprocessing
disposable items may impair the functionality and safety of the
product and lead to unforeseeable reactions as a result of
ageing, embrittlement, wear, thermal load, the effects of
chemical processes, etc.
This product may contain short-term NIBP cuffs. Short-term
NIBP cuffs are only intended for short-term use. They should
thus only be used in the short term and be reprocessed a
maximum of 20 times. With typical use (3 applications per day
followed by disinfection), this results in a recommended total
usage period of 1 week. When reprocessing, note the
information given in the Instructions for Use for the short-term
NIBP cuffs. Reprocessing short-term NIBP cuffs too often may
impair the functionality and safety of the product and lead to
unforeseeable reactions as a result of ageing, embrittlement,
wear, thermal load, the effects of chemical processes, etc.
Wear suitable protective equipment for hygienic reprocessing
work.
Wear suitable gloves for disinfection work.
Please follow the Instructions for Use supplied with the
disinfectant used.
Also follow the respective Instructions for Use for the
components and the accessories.
Always carry out a function check after the hygienic
reprocessing
(see "10.2 Performing a function check", page
148).
You can find further information about hygienic reprocessing
and a list of all the suitable cleaning agents and disinfectants in
a brochure on our website at
www.weinmann-emergency.com.