8 Hygienic reprocessing
Result
8.6 Wipe disinfecting parts
Requirement
184
EN
MEDUMAT Standard
6. Rinse the pressure measuring tube and PEEP control tube with
drinking-water quality from the inside at least eight times with
a disposable syringe.
When doing so, note: Only rinse in one direction.
7. Allow the reusable measuring circuit to dry out completely.
8. Check the reusable measuring circuit for residue and any
remaining dirt.
9. If visible dirt remains: Repeat the manual cleaning.
The reusable measuring circuit has been cleaned manually.
The parts have been manually cleaned and are visibly clean
(see "8.5 Cleaning parts manually", page
1. The parts approved for wipe disinfection can be found in the
cleaning and disinfection plan
disinfection plan", page
2. The cleaning agents, doses and exposure time for the individual
parts can be found in the cleaning and disinfection plan
(see "8.11 Cleaning and disinfection plan", page
3. Wipe-disinfect the parts with one of the agents listed
(see "8.11 Cleaning and disinfection plan", page
When doing so, note:
•
Wipe down all surfaces with a cloth at least 8 times.
2
181).
(see "8.11 Cleaning and
196).
196).
196).