8 Hygienic reprocessing
Result
8.4 Wipe disinfecting parts
Requirement
152
EN
LIFE-BASE
5. With cables: Wrap a disposable cloth around the cable and
pull the cable through the cloth.
6. If soiling is still visible: Repeat the manual cleaning.
7. Wipe down the parts with a damp cloth in order to
remove cleaning agent residue.
8. Allow parts to dry fully at room temperature.
The parts are manually cleaned.
•
The hygienic reprocessing is prepared
hygienic reprocessing", page
•
The parts have been manually cleaned and are visibly clean
(see "8.3 Cleaning parts manually", page
1. Refer to the cleaning and disinfection plan for approved parts,
approved agents and dosages
disinfection plan", page
2. Wipe disinfect parts
page
155).
When doing so, note:
•
All surfaces must be wetted with disinfectant solution.
(see "8.2 Preparing
150).
(see "8.7 Cleaning and
155).
(see "8.7 Cleaning and disinfection plan",
151).