7.5 Disinfecting parts by wiping

Permitted parts
Requirements
The parts have been cleaned manually and are visibly clean
"7.4 Cleaning parts manually", page
1. For parts approved for disinfection by wiping, refer to the
2. For the cleaning agents, dose and exposure time for the
3. Prepare the solution for disinfection as specified by the
4. Disinfect parts by wiping with one of the products listed
Part
Device
Testing bag
12 V cable
Charging adapter
Filter compartment
cleaning and disinfection plan
disinfection plan", page
individual parts, refer to the cleaning and disinfection plan
"7.11 Cleaning and disinfection plan", page
disinfectant manufacturer.
"7.11 Cleaning and disinfection plan", page
Please note:
Wet uneven surfaces and grooves (e.g. navigation knob,
ventilation hose connection) adequately with disinfectant.
When replacing the filter: Disinfect the filter compartment
by wiping.
7 Hygienic reprocessing
Disinfection by wiping
Wipe down with Incidin™ Oxywipe S (Ecolab)
When replacing the filter:
Wipe down with Incidin™ Oxywipe S (Ecolab)
79).
(see "7.11 Cleaning and
92).
MEDUVENT Standard
(see
(see
92).
(see
92).
EN
83