Divert Valve

Divert/Inject Valve Controls

Thermo Scientific
In the divert valve configuration, the valve switches between these two positions:
Detector (position 1–2)—Solvent flow from the LC pump enters the valve through port 5
and exits through port 6 into the API source.
Waste (position 1–6)—Solvent flow from the LC pump enters the valve through port 5 and
exits through port 4 to waste.

Figure 67. Divert/Inject valve plumbed as a divert valve

2
2
3
4
3
No.
Description
1.
Port 1 to API Source
2.
Port 2 from LC Pumps
3.
Port 3 to Waste
You can control the divert/inject valve by using the MS's data system or the control buttons
on the valve.
• Use the MS's data system to specify the parameters in the Divert Valve Properties pane of
the Method Editor. For instructions, refer to the Xcalibur Method Editor Help.
• Use the valve's control buttons
waste when the valve is in the divert valve configuration, or switch between load and
inject modes when the valve is in the loop injector configuration. For instructions, refer to
the manufacturer's manual.
1
1
6
5
(Figure
68) to divert the LC flow between the MS and
TSQ Altis, Quantis, and Fortis Hardware Manual
C
Divert/Inject Valve
Divert/Inject Valve Controls
131