Abort (Remote Command Only)
Stops any sweep or measurement in progress and resets the sweep and trigger system. A
measurement refers to any of the measurements found in the MEASURE menu.
Mode
Remote Command
Example
Notes
Single
Enables single measurements. Pressing Single is equivalent to sending the remote
command sequence:
INIT:CONT OFF
INIT:IMMediate
Single
Therefore,
turns
Mode
Remote Command
Example
254
[:SENSe]:SWEep:POINts <integer>
[:SENSe]:SWEep:POINts?
SWE:POIN 1001
401
2 to 1001
Control/Sweep, Points
all

:ABORt

ABOR
If you are in single sweep/measure (INITiate:CONTinuous is
off), then you must start another sweep (send
INITiate:IMMediate). When the trigger conditions are met, a
new measurement will start.
If you are in continuous sweep/measure
(INITiate:CONTinuous is on), a new measurement begins as
soon as the trigger conditions are met.
Cont
(continuous sweep) off. The current sweep will finish.
SA
:INITiate[:IMMediate]
INIT
Chapter 4