Initial S/W Revision
Compatibility Model
Sets the response to the *IDN command to be a string describing the emulated instrument, such as
"FSP".
Key Path
Example
Initial S/W Revision
System IDN Response
Sets the response to the *IDN command to be a string describing the analyzer model, such as "N9020A".
Key Path
Example
Initial S/W Revision
User IDN Response
Allows you to set or query the user-defined mode IDN response.
This setting affects the response string of "*IDN?", and the whole SCPI tree, according to the current
selection.
The query returns the current User IDN Response string, unless the string is empty, in which case the
System IDN Response is returned instead.
Key Path
Remote Command
Example
State Saved
Initial S/W Revision
ABORt Subsystem
The ABORt subsystem contains commands for aborting triggered actions. An action may be triggered
again immediately after being aborted. All commands trigger events, and therefore have no *RST value.
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
A.06.00 or later
Mode Setup, Mode IDN Response
RLC:IDN:TYPE COMP
A.06.00
Mode Setup, Mode IDN Response
RLC:IDN:TYPE SYST
A.06.00
Mode Setup
[:SENSe]:RLC:IDN:USER <string>
[:SENSe]:RLC:IDN:USER?
SENS:RLC:IDN:USER "abc"
RLC:IDN:USER?
Saved in instrument state.
A.06.00
Compatible SCPI Commands
Supported SCPI Commands
59