Mode
Meas Off (Spectrum Analyzer) Measurement
the measurement to logarithmic (video) averaging, or power (RMS) averaging. Trace
averaging is turned on by pressing
Mode
Remote Command
Example
Preset
State Saved
Key Path
Dependencies /
Couplings
Log-Pwr Avg (Video) Sets the logarithmic (decibel) scale for all filtering and averaging
processes. This scale is also referred to as "Video" because it is the most common display
and analysis scale for the video signal within a spectrum analyzer.
Example
Key Path
Pwr Avg (RMS) Sets the signal power level (the square of the magnitude) as the scale for
the filtering and averaging processes. This scale is sometimes called RMS because the
resulting voltage is proportional to the square root of the mean of the square of the voltage.
Example
Key Path
Limit Masks
Access the Limit Masks submenu keys, enabling you to select to configure either Limit
Mask 1 (Upper) or Limit Mask 2 (Lower).
60
Trace/Detector
Average
,
SA

[:SENSe]:AVERage:TYPE RMS|LOG

[:SENSe]:AVERage:TYPE?

AVERage:TYPE RMS|LOG
AVERage:TYPE?
Log-Pwr Avg (Video)
Saved in instrument state.
Meas Setup, Avg Type
If the Average detector is currently selected then
(Video)
trace averaging is not available (grayed out). You must
Trace/Detector
change the detector (see
Log-Pwr Avg (Video)
type of trace averaging.
AVER:TYPE LOG
Meas Setup, Avg Type, Log-Pwr Avg (Video)
AVER:TYPE RMS
Meas Setup, Avg Type, Pwr Avg (RMS)
.
Log-Pwr Avg
Detector
,
) to access the