Agilent Technologies N1996A-503 User Reference page 164

Csa spectrum analyzer
Table of Contents

Advertisement

Mode
Two Port Insertion Loss Measurement
Mode
Range
Key Path
Avg Mode
Averaging functions are the same as defined in the Spectrum Analyzer Mode, but the
parameter values are measurement independent. So, SCPI commands are different. See
descriptions below.
Avg Mode
sets the type of termination control used for the averaging function. This
determines the averaging action after the specified Averaging number has been reached.
Mode
Remote Command
Example
Preset
Range
Key Path
164
Stimulus / Response
Averaging | Average | Clear Normalize | Limits | Optimize |
Interference Rejection | Store As
Front-panel key
Stimulus / Response
[:SENSe]:TPILoss:AVERage:TCONtrol
EXPonential|MAXHold|REPeat|RMAXhold
[:SENSe]:TPILoss:AVERage:TCONtrol?
[:SENSe]:TPILoss:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:TPILoss:AVERage[:STATe]?
TPIL:AVER:TCON RMAX
TPIL:AVER ON
EXPonential, ON
EXPonential|MAXHold|REPeat|RMAXhold
Meas Setup, Avg Mode
Chapter 3