Mode
Two Port Insertion Loss Measurement
Avg Limits On/Off Turns on/off the average limit values.
Mode
Preset
Range
Key Path
Max Loss Limit Sets the maximum average limit value. If the average limits are turned on, then
the measured average insertion loss will be compared with this value.
Mode
Preset
Key Path
Min Loss Limit Sets the minimum average limit value. If the average limits are turned on, then
the measured average insertion loss will be compared with this value.
Mode
Preset
Key Path
Optimize
When using the default settings, the spectrum analyzer optimizes measurement accuracy
at the expense of dynamic range. There may be times, however, when you prefer to gain
dynamic range and lose some accuracy. For example, you may be making a measurement
where you want to see a relatively large dynamic range, and your results will not be
impaired significantly by 1 to 2 dB less accuracy. Optimizing for dynamic range gains you
approximately 30 dB of dynamic range.
Mode
Remote Command
Notes
Preset
176
Stimulus / Response
0
On|Off
Meas Setup, Avg Limits, Avg Limits On/Off
Stimulus / Response
20 dB
Meas Setup, Avg Limits, Max Loss Limit
Stimulus / Response
50 dB
Meas Setup, Avg Limits, Min Loss Limit
Stimulus / Response
N/A. Front Panel feature only
Accy
Chapter 3