Mode
Meas Off (Spectrum Analyzer) Measurement
Notes
Dependencies / Couplings
Preset
Range
Key Path
Ext Gain
Sets a correction value to compensate for gain (or loss) through an external amplifier or
attenuator connected between a signal source and the RF input. The correction value is
applied to the displayed measurement values.
Mode
Remote Command
Notes
Preset
Min
Max
Key Path

Auto Tune

Auto Tune
is an immediate action key. When pressed,
change center frequency to center the strongest signal on screen. When
decreases the span significantly, the signal may not be centered due to the increased
36
If the preamplifier option (Option P03 or P06) is not installed
or is unlicensed, the key is grayed out and displays OFF as a
state.
When Auto Range is On, the state of this Preamp is
automatically determined by the measurement.
When the user sets a value manually, Auto Range
automatically changes to Off.
Per default, the "Internal Preamp" is in automatic mode. See
section 1.2.5.2.1 Auto Range
OFF|ON|0|1
AMPTD Y Scale, More (2 of 2), Internal Preamp
SA
[:SENSe]:CORRection:OFFSet[:MAGNitude] <rel_ampl>
[:SENSe]:CORRection:OFFSet[:MAGNitude]?
[:SENSe]:CORRection:OFFSet[:MAGNitude]:STATe
ON|OFF|1|0
[:SENSe]:CORRection:OFFSet[:MAGNitude]:STATe?
The setting of this "External Gain" parameter will be preserved
through a Mode Preset but not through a User Preset or Recall
<state>.
ON
-100 dB
100 dB
AMPTD Y Scale, More (2 of 2), Ext Gain
Auto Tune
causes the analyzer to
Auto Tune
Chapter 3