Agilent Technologies N1996A-503 User Reference page 217

Csa spectrum analyzer
Table of Contents

Advertisement

the
Segments
key.
Mode
Notes
Dependencies /
Couplings
Preset
State Saved
Range
Key Path
Level: 2nd Segment Left
Sets the power level, for the 2
Segments
has been selected with the
Mode
Notes
Preset
Chapter 3
Stimulus / Response
Key is only displayed when the selected segment is 2
The span for each of the left
The Left span is calculated as follows:
(<maximum span>/2) – (<center segment span>/2) - <first
segment left span>
The Right span is calculated as follows:
(<maximum span>/2) – (<center segment span>/2) - <first
segment right span>
Where:
<maximum span> is the maximum frequency span of the
instrument (3 GHz or 6 GHz)
<center segment span>
<first segment left span>
<first segment right span>
Not affected by a preset but set per default to Left after a power
cycling
Saved in instrument state.
Left | Right
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment (Left or Right
Limits)
nd
segment to the Left of the Center segment. This key is only available when
key.
Stimulus / Response
Key is only displayed when the selected segment is 2
selected segment side is Left.
Not affected by a preset but set per default to 30 dB after a
power cycling
Mode
Return Loss
nd
.
nd
2
nd
, and the
217

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1996a-506

Table of Contents