Calculate:lmask[1]|2:Second:segment:left:level <Ampl; Calculate:lmask[1]|2:Second:segment:left:level - Agilent Technologies N1996A-503 User Reference

Csa spectrum analyzer
Table of Contents

Advertisement

Dependencies /
Couplings
Remote Command
Notes
Preset
State Saved
Range
Key Path
Level: 2nd Segment Left
Sets the power level, for the 2
nd
2
available when
has been selected with the
Mode
Remote Command
Example
Notes
Preset
State Saved
Key Path
Chapter 3
The Left span is calculated as follows:
(<maximum span>/2) – (<center segment span>/2) - <first
segment left span>
The Right span is calculated as follows:
(<maximum span>/2) – (<center segment span>/2) - <first
segment right span>
Where <maximum span> is the maximum frequency span of
the instrument (3 GHz or 6 GHz)
The Limit Mask SCPI command syntax uses:
The keyword "LEFT" to define the level and span for the second
segment to the Left of the center segment.
The keyword "RIGHt" to define the level and span for the
second segment to the Right of the center segment.
Not affected by a preset but set per default to Left after a power
cycling.
Saved in instrument state.
Left | Right
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment (Left or Right)
nd
segment to the Left of the Center segment. This key is only
Segments
Spectrum Analyzer

:CALCulate:LMASk[1]|2:SECond:SEGMent:LEFT:LEVel

<ampl>
:CALCulate:LMASk[1]|2:SECond:SEGMent:LEFT:LEVel?
:CALC:LMAS:SEGM:SEC:LEFT:LEV –39 dBm
Key is only displayed when the selected segment is 2
selected segment side is Left.
Not affected by a preset but set per default to -20 dB after a
power cycling
Saved in instrument state.
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment: Left, Level
Meas Off (Spectrum Analyzer) Measurement
key.
Mode
nd
, and the
69