Min
Max
Key Path
Restart
Restart
Pressing
restarts the current sweep, or measurement, or set of averaged/held
sweeps or measurements. The
as quickly as possible. It then resets the sweep and trigger systems, sets up the
measurement and initiates a new data measurement sequence with a new data acquisition
(sweep) taken once the trigger condition is met.
Mode
Remote Command
Key Path
Points
Sets the number of points per sweep, from 2 to 1001. The default setting is 401 points.
Increasing the number of sweep points increases frequency resolution and accuracy but
reduces measurement speed. The current value of points is displayed next to the sweep
time in the lower-right corner of the display.
Changing the number of points has several effects on the analyzer. Since markers are read
at the point location, the marker reading may change. Trace data for all traces is cleared. If
sweep is set to Cont, a new sweep begins immediately. If any trace is in average or hold,
the averaging starts over. If limit masks are set to On, the limit masks are updated.
In addition to sweep points, the span, resolution bandwidth, average detection and center
frequency will also affect measurement speed.
Mode
Chapter 4
Saved in State
2/RBW or 1 µs whichever is higher
102400/RBW or 10 s whichever is lower but will be never lower
than 27.307 ms.
Exact coupling equation controlling the Max Value of the Sweep
Time is:
204800 / (2*RBW) seconds for RBW <= 1.25 MHz
204800 / 7.5e6 = 27.3067 ms for RBW > 1.25 MHz
Control/Sweep, Sweep Time
Restart
function first aborts the current sweep/measurement
SA
:INITiate:RESTart
Control/Sweep, Restart
SA
Utility and Control/Sweep Functions
Control/Sweep Functions
253