Mode
Meas Off (Spectrum Analyzer) Measurement
Preset
State Saved
Key Path
Default Unit
Level: 1st Segment Right
Sets the power level, for the 1
1st
available when
has been selected with the
Mode
Remote Command
Example
Notes
Dependencies /
Couplings
Preset
State Saved
Key Path
Default Unit
Span: 1st Segment Left
Sets the frequency span for the 1
1st
available when
has been selected with the
Mode
Remote Command
Example
66
Not affected by a preset but set per default to –30 dBm after a
power cycling
Saved in instrument state.
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Left, Level
dBm
st
segment to the Right of the Center segment. This key is only
Segments
SA
:CALCulate:MASk[1]|2:FIRSt:SEGMent:RIGHt:LEVel
<ampl>
:CALCulate:MASk[1]|2:FIRSt:SEGMent:RIGHt:LEVel?
:CALC:LMAS:SEGM:FIRS:RIGH:LEV –14 dBm
Key is only displayed when the selected segment is 1
selected segment side is Right.
Value is displayed on the key in the currently selected Y Axis
Unit.
Not affected by a preset but set per default to -10 dBm after a
power cycling
Saved in instrument state.
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Right Level
dBm
st
segment to the Left of the Center segment. This key is only
Segments
SA
:CALCulate:LMASk[1]|2:SEGMent:FIRSt:LEFT:SPAN
<frequency>
:CALCulate:LMASk[1]|2:SEGMent:FIRSt:LEFT:SPAN?
:CALC:LMAS1:SEGM:FIRS:LEFT:SPAN 1.75 MHz
key.
key.
st
and the
Chapter 3