instrument begins a new sweep and the trace is updated.
Example
Dependencies / Couplings
State Saved
Key Path
Average
Average
Activates the
function for the selected trace. Pressing
displays an average trace, which represents the cumulative average on a point-to-point
basis of the new trace data and previous averaged trace data.
Example
Dependencies / Couplings
State Saved
Key Path
Max Hold
Max Hold
Pressing
causes the trace to be cleared and a new maximum-hold measurement
sequence is started. Max hold maintains the maximum level for each trace point of the
selected trace, and updates each trace point if a new maximum level is detected in a
successive sweep. Max hold is restarted if any parameter is changed. Restarting max hold
also restarts averaging and min hold, because there is only one counter for Average and
Hold.
Example
Dependencies / Couplings
Chapter 3
TRAC:TYPE WRIT
Whenever the user presses
, Update
command
is set to
The type information for each trace is saved in Instrument State
Trace/Detector, Clear Write
TRAC:TYPE AVER
Avg Mode
Affected by
and
Whenever the you press
, Update
command
is set to
Automatic detector selection rules depend on the trace type
selections.
The type information for each trace is saved in Instrument State
Trace/Detector, Average
TRAC:TYPE MAXH
Avg Mode
Affected by
and
Whenever the user presses
, Update
SCPI command
is set to
Automatic detector selection and the VBW:RBW ratio auto
rules both depend on the trace type selections.
Meas Off (Spectrum Analyzer) Measurement
Clear Write
or sends the equivalent SCPI
On
Display
and
is set to
Average
maintains and
Avg Number
Average
or send the equivalent SCPI
On
Display
and
is set to
Average/Hold Count
Max Hold
or sends the equivalent
On
Display
and
is set to
Mode
Show
.
Show
.
Show
.
79