Agilent Technologies N1996A-503 User Reference page 354

Csa spectrum analyzer
Table of Contents

Advertisement

IEEE and STATus Commands (Remote Commands Only)
Agilent CSA STATus Register Model
Questionable Negative Transition
This command determines which bits in the Questionable Condition register will set the
corresponding bit in the Questionable Event register when the condition register bit has a
negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the
bits that you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Positive Transition
This command determines which bits in the Questionable Condition register will set the
corresponding bit in the Questionable Event register when the condition register bit has a
positive transition (0 to 1). The variable <integer> is the sum of the decimal values of the
bits that you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
354
All modes
:STATus:QUEStionable:NTRansition <integer>
:STATus:QUEStionable:NTRansition?
STAT:QUES:NTR 32 Frequency summary 'questionable
cleared' will be reported to the Status Byte Register.
0
0
32767
Sequential command
All modes
:STATus:QUEStionable:PTRansition <integer>
:STATus:QUEStionable:PTRansition?
STAT:QUES::PTR 32 Frequency summary 'questionable
cleared' will be reported to the Status Byte Register.
32767
0
32767
Sequential command
Chapter 7