Resolution Distance (in meters) = Measured distance (in meters) / (1/2 * Number of
Points)
Please be careful how you interpret this equation. Note that to increase the resolution, you
need to reduce the Resolution Distance; to reduce the resolution, you need to increase the
Resolution Distance.
You can see from this equation that:
To increase the resolution, that is, to reduce the Resolution Distance:
• you can increase the number of points, or
• you can reduce the measured distance.
To reduce the resolution, that is, to increase the Resolution Distance, you can
• you can reduce the number of points, or
• you can increase the measured distance.
Cable Type
Accesses the features for defining the type of cable that you are measuring.
Mode
Key Path
Cable Type Selects the type of cable. Each time you press this key, the selected type
changes. If you select RG or BTS, you will be asked to select the specific type of cable you
are measuring from a list of common cable types.
If the type of cable you are measuring is not listed, you need to select
Mode
Range
Key Path
Select Cable Each time you press this key, the selected option changes. If you select RG or
BTS, you will be asked to select the specific type of cable you are measuring from a list of
common cable types.
Chapter 3
Stimulus / Response
Meas Setup, Cable Type
Stimulus / Response
RG | BTS | Cust
Meas Setup, Cable Type
Mode
Distance To Fault
Cust
(Custom Cable).