Agilent Technologies N1996A-503 User Reference page 158

Csa spectrum analyzer
Table of Contents

Advertisement

Mode
Two Port Insertion Loss Measurement
Example
Dependencies /
Couplings
Preset
State Saved
Range
Key Path
Stop Freq
Use this key to specify the Stop Frequency of your frequency range.
Mode
Remote Command
Example
Dependencies /
Couplings
Preset
State Saved
Range
Key Path
Normalize
Before you perform a two port insertion loss measurement, you must first normalize the
measured values for insertion loss by compensating for the loss associated with the devices
(adapters, cables) that connect the Agilent CSA Spectrum Analyzer to the device or
assembly being tested. Otherwise, the loss introduced by these connecting devices is added
to the loss of the device under test.
158
FREQ:STAR 200 MHz
FREQ:STAR?
You cannot set the Start frequency equal to the Stop frequency.
The instrument will alter the value of the last setting to
maintain a minimum value of 1 kHz difference between the
Start and Stop frequencies.
1.85 GHz
Saved in instrument state.
From 10 MHz to 3 GHz (Option 503) / 6 GHz (Option 506)
FREQ/Channel, Start Freq
Stimulus / Response
[:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
FREQ:STOP 220 MHz
FREQ:STOP?
You cannot set the Start frequency equal to the Stop frequency.
The instrument will alter the value of the last setting to
maintain a minimum value of 1 kHz difference between the
Start and Stop frequencies.
1.99 GHz
Saved in instrument state.
From 10.001 MHz to 3 GHz (Option 503) / 6 GHz (Option 506)
FREQ/Channel, Stop Freq
Chapter 3