Frame Skip Enables you to set the number of frames you would like to skip when capturing
data. You may set this value from 0 to 2,147,483,647 frames. Increasing the frame skip
value causes the display to redraw the spectrum every nth frame and a block of lines are
shown at once instead of a single line at a time. Higher frame skip values are for use with
fast measurements.
Mode
Preset
State Saved
Min
Max
Key Path
Palette Switches the spectrogram color mode between full color and grayscale.
Mode
Preset
Range
Key Path
Display Enable (Remote Command Only)
Turns the display on/off, including the display drive circuitry and the backlight. The three
goals of turning the display off are:
• To increase speed as much as possible by freeing the instrument from having to update
the display
• To reduce emissions from the display, drive circuitry and backlight
Chapter 3
Max Speed (0 s.)
Saved in State
400 nS
(200*1024)/(2*RBW)
View/Display, Spectrogram, Update Interval
SA
0 frames
Saved in State
0
2,147,483,647
View/Display, Spectrogram, Frame Skip
SA
Full Color
Full Color | Grayscale
View/Display, Spectrogram, Palette