Agilent Technologies N1996A-503 User Reference page 243

Csa spectrum analyzer
Table of Contents

Advertisement

Mode
Notes
Preset
State Saved
Min
Max
Key Path
Default Unit
Level: 2nd Segment Right
Sets the power level, for the 2
nd
2
available when
has been selected with the
Mode
Notes
Preset
State Saved
Min
Max
Key Path
Default Unit
FFT Size
This sets the FFT size by specifying the number of FFT points.
NOTE
You can set the number of points to 256, 512, or 1024, but you will only
be able to save 256 data points when you save trace data. This is
because only 256 points are ever used to display the trace, regardless of
how many points you have used to actually make the measurement.
You will not, however, be losing any resolution, or reducing the quality
Chapter 3
Stimulus / Response
Key is only displayed when the selected segment is 2
selected segment side is Left.
Not affected by a preset
Saved in instrument state.
–10 dB
150 dB
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment: Left Level
dB
nd
segment to the Right of the Center segment. This key is only
Segments
Stimulus / Response
Key is only displayed when the selected segment is 2
selected segment side is Right.
Not affected by a preset
Saved in instrument state.
–10 dB
150 dB
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment: Right Level
dB
key.
Mode
Distance To Fault
nd
, and the
nd
, and the
243