Agilent Technologies
N9010/20/30AK-REA
Installation Guide
Refer to Agilent X Series Signal Analyzers
Option SSD Solid State Disk Drive
Installation Note (N9030-90009)
Manufacturing Part Number: N9030-90019
Printed in USA: July 2011
Supersede: March 2010
© Copyright Agilent Technologies, Inc. 2010, 2011

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies N9010AK-REA

  • Page 1 Agilent Technologies N9010/20/30AK-REA Installation Guide Refer to Agilent X Series Signal Analyzers Option SSD Solid State Disk Drive Installation Note (N9030-90009) Manufacturing Part Number: N9030-90019 Printed in USA: July 2011 Supersede: March 2010 © Copyright Agilent Technologies, Inc. 2010, 2011...
  • Page 2 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
  • Page 3 Safety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document. WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life.
  • Page 4 Installation Guide...
  • Page 5: Table Of Contents

    Contents N9010/20/30AK-REA Description ..............2 Verifying the Shipment .
  • Page 6 Contents Contents-2...
  • Page 7: N9010/20/30Ak-Rea

    N9010/20/30AK-REA N9030-90019...
  • Page 8: Description

    N9010/20/30AK-REA Description Description The Agilent N9010/20/30AK Option REA provides the capability to relocate the instrument disk drive from the internal drive drawer, to an external customer supplied enclosure. This is accomplished by installing an eSATA drive adapter into the CPU rear panel slot, normally occupied by the internal disk drive.
  • Page 9: Verifying The Shipment

    Inspect the shipping container. If the container or packing material is damaged, it should be kept until the contents of the shipment have been checked mechanically and electrically. If there is physical damage please notify the nearest Agilent Technologies office. Refer to “Agilent Support, Services, and Assistance” on page 12.
  • Page 10: Hard Drive Disk Or Solid State Drive Replacement Procedure

    N9010/20/30AK-REA Hard Drive Disk or Solid State Drive Replacement Procedure Hard Drive Disk or Solid State Drive Replacement Procedure The hard disk drive and solid state disk drive will be referred to as HDD/SSD through out this document. Follow the appropriate Electrostatic Discharge (ESD) procedures. ESD can CAUTION damage or destroy electronic components.
  • Page 11: Return For Service

    N9010/20/30AK-REA Hard Drive Disk or Solid State Drive Replacement Procedure 3. Remove the 2.5 inch HDD/SSD from its tray assembly by removing the 4 screws. NOTE The tray assembly must be kept for use when returning the instrument for calibration or repair. 4.
  • Page 12: Safety And Regulatory Information

    Declaration of Conformity A copy of the Declaration of Conformity is available upon request, or a copy is available on the Agilent Technologies web site at http://regulations.corporate.agilent.com/DoC/search.htm Statement of Compliance This instrument has been designed and tested in accordance with CAN/CSA 22.2 No.
  • Page 13: Before Applying Power

    N9010/20/30AK-REA Safety and Regulatory Information Before Applying Power Verify that the premises electrical supply is within the range of the instrument. The instrument has an autoranging power supply. WARNING If this product is not used as specified, the protection provided by the equipment could be impaired.
  • Page 14: Servicing

    Allow all residual alcohol moisture to evaporate, and fumes to dissipate prior to energizing the instrument. To prevent electrical shock, disconnect the WARNING Agilent Technologies from mains electrical supply before cleaning. Use N9010/20/30AK-REA a dry cloth or one slightly dampened with water to clean the external case parts.
  • Page 15: Electrostatic Discharge Protection

    N9010/20/30AK-REA Electrostatic Discharge Protection Electrostatic Discharge Protection Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables to the instrument. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage.
  • Page 16: Regulatory Information

    N9010/20/30AK-REA Regulatory Information Regulatory Information This section contains information that is required by various government regulatory agencies. Instrument Markings The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation. This symbol indicates that the instrument requires alternating current (ac) input.
  • Page 17: Battery Collection

    Regulatory Information Battery Collection Do not throw batteries away but collect as small chemical waste, or in accordance with your country’s requirements. You may return the battery to Agilent Technologies for disposal. Refer to “Contacting Agilent” on page 12 for assistance.
  • Page 18: Agilent Support, Services, And Assistance

    Agilent for repair. If you wish to send your instrument to Agilent Technologies for service or repair: • Include a complete description of the service requested or of the failure and a description of any failed test and any error message.