ABORt Subsystem

ABORt Subsystem

Abort

:ABORt

Restarts any sweep or measurement in progress and resets the sweep
or trigger system. A measurement refers to any of the measurements
found in the
MEASURE
If :INITiate:CONTinuous is off (single measure), then
:INITiate:IMMediate will start a new single measurement.
If :INITiate:CONTinuous is on (continuous measure), a new
continuous measurement begins immediately.
The INITiate and TRIGger subsystems contain additional related
commands.
Front Panel
Access:
Restart
5-10
menu.
for continuous measurement mode