Calculate:data - Agilent Technologies N1996A-503 User Reference

Csa spectrum analyzer
Table of Contents

Advertisement

06,2.14748e+06,2.14748e+06,2.14748e+06,2.14748e+06,2.14748e+06,2.14748e+06,2.14748
e+06,2.14748e+06,2.14748e+06,2.14748e+06,2.14748e+06,2.14748e+06,2.14748e+06,2.147
48e+06,2.14748e+06,2.14748e+06,2.14748e+06,0,1,0,1,0
This response is in standard block format. The above example (#45732) shows that there
are 5732 data bytes in the results string. Since this string is in ASCII, 1 data byte is
actually 1 character in the string. For example, "4.076e+00" are 9 bytes.
The response will always begin with a #, then the next number will represent the number
of digits (i.e. 4) of the "data bytes" number in the results string. This will then indicate that
the next 4 digits will be the block size, 5732 data bytes.
After this header information, you get the data in a comma separated list.
Mode
Remote Command
Example
Notes
Remote Command Notes
Key Path
AMPTD/Y Scale
AMPTD/Y Scale
The following
measurement. The measurement must already be running to access these keys. Press
Meas
Two Port Insertion Loss
,
Mode
Key Path
Chapter 3
Stimulus Response

:CALCulate:DATA?

CONF:TPIL selects the command.
INIT starts a measurement cycle.
*OPC? holds off the next command until the above
commands are complete
:CALC:DATA?
You must be in the Stimulus Response mode and in the
TPIL measurement in order to get the TPIL results using
this CALCulate:DATA? query.
Agilent CSA platform does not support the READ, FETCH,
MEAS commands but offers this CALCulate:DATA?
command to get the data results of the active measurement.
N/A. Function is only available remotely., Two Port Insertion
Loss
keys are for setting up a Two Port Insertion Loss
to turn on the measurement.
SR
Front-panel key
Two Port Insertion Loss Measurement
Mode
155