Inverse Time Sets the Marker Readout (Marker X Axis scale) to Inverse Time, displaying
the reciprocal of the marker delta time. It is useful in a delta mode to show the reciprocal
of (sweep) time between two markers. This function is only meaningful when on a time
domain trace and in the Delta control mode.
Sets the marker readout to inverse time. In non-zero span, this is 1/frequency. This key is
not available in Meas Off mode.
Mode
Remote Command
SCPI Example
Remote Command Notes
Dependencies /
Couplings
Key Path
Marker All Off
Turns off all the markers on all traces.
Mode
Remote Command
Remote Command Notes
Key Path
Marker X Axis Value (Remote Command Only)
Sets the marker X Axis value in the current marker X Axis Scale unit. It has no effect if
the control mode is Off, but is the SCPI equivalent of entering an X value if the control
Chapter 3
Spectrum Analyzer with Meas OFF
See details in "Marker Readout (Marker X Axis Scale Type)"
Table described above
CALC:MARK2:X:READ ITIM sets marker 2 with a Marker
Readout type (Marker X Axis scale) to Inverse Time.
Inverse Time" parameter is not available in the Frequency
Domain, and in the Time Domain, the parameter is only
accessible if Marker Mode is set to Delta. Note the following
equation:
Inverse Time Marker Readout = 1 / Time Delta Marker Readout
If the Normal and Delta markers are at the same X Axis value,
the time between them is 0, so the reciprocal of sweep time is
infinitely large. The display will show "---" and a SCPI query
will return not a number (NAN).
Only available when x axis scale is time and a delta marker is
selected.
Marker, Marker Readout, Inverse Time
All

:CALCulate:MARKer:AOFF

CALC:MARK:AOFF
Marker, Marker All Off
Meas Off (Spectrum Analyzer) Measurement
Mode
51