Agilent Technologies N1996A-503 User Reference page 258

Csa spectrum analyzer
Table of Contents

Advertisement

Utility and Control/Sweep Functions
Utility Functions
Remote Command Notes
Key Path
Type
Enables you to select the type of file you want to recall. The file types available for recall
are described below.
Type
State
Trace
Mode
Preset
State Saved
Key Path
258
For the MMEM:LOAD:TRAC <trace>,<'file_name'> command,
<'file_name'> must include the following:
• A file extension of TRC. The .trc extension is for files that
include both trace and state data.
• File names are case sensitive.
• Trace names are:
TRACE1|TRACE2|TRACE3|TRACE4|ALL
Measurements in the Stimulus/Response mode have only
TRACE1|TRACE2 available
For the MMEM:LOAD:STAT 1 command:
• This command only recalls the state of the Signal Analysis
mode. Use *RCL to recall all mode states.
*RCL <register#> recalls the state from the specified internal
register into the current active state of the instrument. The
available register numbers are 0 to 127
When you are making measurements in the Stimulus/Response
Mode, the :MMEMory:LOAD:STATe 1,<'file_name'> command is
only supported in the Two-Port Insertion Loss measurement. It
is not supported in the One-Port Insertion Loss, Return Loss, or
Distance To Fault measurements.
Recall, Recall Now
Format
State
Trace + state
All
Type survives Preset and *RST, but is set to State at power on.
Type is not saved in the instrument state.
Recall, Type
Source
Extension
sta
Trace 1, 2, 3, 4, or all
trc
traces
Chapter 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1996a-506