Agilent Technologies N1996A-503 User Reference page 233

Csa spectrum analyzer
Table of Contents

Advertisement

distance to measured distance times the number of points you have specified. The higher
the number of data points, the greater the measurement resolution.
In most cases, the default resolution using 256 data points will be adequate to locate the
faults, but if more resolution is needed you can increase the span between the start and
stop frequencies (which will decrease the measured distance) or use the other approach,
automatic frequency range. If the measurement distance is not long enough for the cable
you are testing, reduce the span between the start and stop frequencies (which will
increase the measurement distance) or use automatic frequency range.
When testing cables attached to antennas, test signals are radiated from the test antenna
Verify that the signal used for the test, and therefore being radiated from the test antenna,
cannot interfere with other radiated signals from other antennas.
To configure measurement settings, select
Meas Setup
This section describes the keys available with
Meas Setup
keys used to set up a Distance to Fault measurement. That measurement must
already be selected by pressing
Mode
Key Path
Avg Mode
Sets the type of termination control used for the averaging function. This determines the
type of averaging used for the specified number of measurements.
Mode
Preset
Range
Key Path
Off Turns averaging functions off.
Key Path:
Chapter 3
Distance to Fault
Distance to Fault
Meas
Distance to Fault
,
Stimulus / Response
Front-panel key
Stimulus / Response
Exponential, On
Off | Exponential | Repeat | Repeat Max Hold
Meas Setup, Avg Mode
Meas setup, Avg Mode, Off
Meas Setup
and then press
measurement. To access the
.
Mode
Distance To Fault
.
233