Mode
Adjacent Channel Power—ACP (I&M) Measurement
FREQ Channel
Accesses the frequency/channel settings that can be changed when using the ACP
measurement.
Mode:
Key Path
FREQ / Channel
FREQ / Channel
The
front-panel key displays the frequency or channel menu functions
available with the adjacent channel power measurement. The frequency/channel keys can
Units
be toggled with the
Channel
Sets the channel value and the channel type:
Forward (Fwd or Dn): the base station is transmitting (downlink)
Reverse (Rev or Up): the mobile is Transmitting (uplink)
NOTE
Dn & Up are used for UARFCN UMTS, UARFCN Band B UMTS, and
all GSM/EDGE channel standards. All other channel standards use
Fwd & Rev.
Mode
Dependencies /
Couplings
Preset
State Saved
Range
Key Path
132
SA
Front-panel key
menu key.
Channel Analyzer
The "Channel" parameter is coupled with the "Center
Frequency" parameter and is dependant on the Channel
Standard selected for the measurement. For instance, if the
"Channel Standard" selected is GSM–850, if the "Channel" is
set to 128 Fwd, the "Center Frequency" value is automatically
changed to 869.2 MHz. Conversely, if the "Center Frequency"
parameter is set to 893.8 MHz, then the "Channel" value is
automatically changed to 251 Fwd.
500 Fwd
Saved in instrument state.
Dependant on the Channel Standard selected for the
measurement.
FREQ/Channel, Channel
Chapter 3