One Port Insertion Loss Measurement
One Port Insertion Loss Measurement
One Port Insertion Loss measurement Description
The one port insertion loss measurement allows you to quantify signal loss in a cable or
other device without connecting both ends of the cable or device to the test set. This
measurement can be especially useful in measuring the loss of a feedline connected to the
antenna on a tower. This method of measuring antenna feedline insertion loss is accurate
for results up to 10 dB.
This measurement is less accurate than Two Port Insertion Loss. When it is not practical
to connect both ends of a device to the test set or for insertion loss measurements greater
than 10 dB — for example when measuring a 40 dB high power attenuator — it is better to
use Two Port Insertion Loss.
To configure measurement settings, select
Setup
Control/Sweep
. Pressing
refer to
"Control/Sweep" on page
NOTE
The test signal emitted from the RF Output connector is a sine wave,
swept from the starting frequency to the stop frequency. When testing
cables that are attached to antennas, test signals are radiated, and
these test signals can cause interference. Verify that the signal used for
the test cannot cause interference to another antenna.
Mode:
Key Path
AMPTD/Y Scale
AMPTD/Y Scale
The following
measurement. The measurement must already be running to access these keys. Press
Meas
One Port Insertion Loss
,
Autoscale
Executes an automatic scaling of the
function and has no SCPI command.
Mode
Key Path
182
One Port Insertion Loss
enables you to restart a measurement, for more information,
98.
Stimulus/Response
Stimulus/Response
keys are for setting up a One Port Insertion Loss
to turn on the measurement.
Reference Level
SA
Front-panel key
and then press
setting. Autoscale is a GUI only
Meas
Chapter 3

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1996a-506

Table of Contents