Mode
Distance To Fault
If the type of cable you are measuring is not listed, you need to select
Mode
Key Path
Cable Atten Use this key to specify the attenuation per unit distance of the cable you are
measuring. You only need to specify this parameter if the cable you are testing is not
shown in the list of common cable types.
If the type of cable you are measuring is not listed, you need to select
and then specify its attenuation in dB per 100 m (or in dB per 100 ft.) You will also need to
specify the speed at which signals travel along this cable [Vel Factor], measured as a
percentage of the speed of light in a vacuum.
Most cable manufactures provide a table showing attenuation versus frequency for each
type of cable. Remember that attenuation varies with frequency, so select an attenuation
value that corresponds as closely as possible to the center of the frequency range you are
using.
Mode
Key Path
Vel Factor Use this key to specify the relative propagation velocity of the cable you are
measuring. You only need to specify this parameter if the cable you are testing is not
shown in the list of common cable types.
If the type of cable you are measuring is not listed, you need to select
and then specify the speed at which signals travel along this cable measured as a
percentage of the speed of light in a vacuum. You will also need to specify its attenuation in
dB per 100 m (or in dB per 100 ft.) Most cable manufactures provide a table showing
attenuation versus frequency for each type of cable.
Mode
Range
Key Path
246
Stimulus / Response
Meas Setup, Cable Type, Select Cable
Stimulus / Response
Meas Setup, Cable Type, Cable Atten
Stimulus / Response
0.0 to 100.0 (%C – speed of light)
Meas Setup, Cable Type, Vel Factor
Cust
(Custom Cable).
Cust
(Custom Cable),
Cust
(Custom Cable),