Limit Masks Sets the Limit Masks submenu keys, it is then possible to select to configure
either Limit Mask 1 (Upper) or Limit Mask 2 (Lower).
Mode
Dependencies /
Couplings
State Saved
Key Path
Limit Mask 1 (upper) or Limit Mask 2 (Lower)
Limit Mask 1 (upper) access the upper limit mask keys.
Limit Mask 2 (Lower) access the lower limit mask keys.
You can define a limit mask that matches your maximum or minimum measurement requirements.
This mask can then be compared with the measured trace data to determine pass/fail conditions.
Limit Trace
Selects the trace to which the currently selected limit mask (1 or 2) will be applied. The limit masks can be
applied to separate traces. Only 2 traces are available in the Stimulus Response mode (compared to 4 traces in
the Spectrum Analyzer Mode).
Mode
Dependencies /
Couplings
Preset
State Saved
Range
Key Path
Chapter 3
Stimulus / Response
The Limit Mask 1 (Upper) menu link key accesses the keys to
configure the upper limit mask for the Center, 1
segments (dependent on selected segment).
The Limit Mask 2 (Lower) menu link key accesses the keys to
configure the lower limit mask for the Center, 1
segments (dependent on selected segment).
Saved in instrument state.
Meas Setup, Limits, Limit Masks
Stimulus / Response
The limit mask is always displayed when the center segment is
turned on; even if the trace to which it is being applied is
blanked ('Show' set to Off).
1
Saved in instrument state.
1 | 2
Meas Setup, Limits, Limit Masks, Limit Mask 1, Limit Trace
or
Meas Setup, Limits, Limit Masks, Limit Mask 2, Limit Trace
Two Port Insertion Loss Measurement
st
nd
, or 2
st
nd
, or 2
Mode
167