Repeat Sets the averaging function to the repeat function. Repeat averaging shows
resulting data after the selected number of sweeps has been averaged, then clears the data
before repeating the process.
NOTE
While Repeat averaging is running, the results from the last completed
measurement remains on the display, the results are not updated until
the last average count has been reached.
Example
Key Path
Repeat Max Hold Sets the averaging function to repeat with max hold function. Repeat
with max hold captures and holds the highest magnitude data at the end of each sweep
count.
NOTE
While Repeat Max Hold is running, the results from the last completed
measurement remains on the display, the results are not updated until
the last average count has been reached. (Maximum data points are
held with each consecutive sweep.)
Example
Key Path
Avg Number
Sets the number of measurement averages used when calculating the measurement result.
Mode
Remote Command
Example
Chapter 3
Adjacent Channel Power—ACP (I&M) Measurement
Meas setup, Avg Mode, Max Hold
ACP:AVER:TCON REP
Meas setup, Avg Mode, Repeat
ACP:AVER:TCON RMAX
Meas setup, Avg Mode, Repeat Max Hold
Channel Analyzer
[:SENSe]:ACPower:AVERage:COUNt <integer>

[:SENSe]:ACPower:AVERage:COUNt?

AVER:COUN 100
Mode
137