Agilent Technologies N1996A-503 User Reference page 142

Csa spectrum analyzer
Table of Contents

Advertisement

Mode
Adjacent Channel Power—ACP (I&M) Measurement
Channel BW/Time/Freq Err Opens a menu of channel bandwidth and time settings that
can be changed when the
Mode
Dependencies /
Couplings
State Saved
Key Path
Center Chan Meas BW Sets the center channel measurement bandwidth.
Mode
Dependencies /
Couplings
Preset
State Saved
Min
Max
Key Path
Center Chan Meas Time Sets the center channel measurement time.
Mode
Dependencies /
Couplings
Preset
State Saved
Min
Max
Key Path
142
Format Type
Cust
is set to
Channel Analyzer
Format Type must be set to List
Saved in State
Meas Setup, Format/BW, Channel BW/Time/Freq Err
Channel Analyzer
Format Type must be set to Cust
200 kHz
Saved in instrument state.
10 kHz
5 MHz
Meas Setup, Format/BW, Channel BW/Time/Freq Err, Center Chan
Meas BW
Channel Analyzer
Format Type must be set to Cust
23.077 ms
Saved in instrument state.
BW dependent
BW dependent
Meas Setup, Format/BW, Channel BW/Time/Freq Err, Center Chan
Meas Time
.
Chapter 3