Preset
State Saved
Min
Max
Key Path
Default Unit
1st Segment
Sets the state of the first segment to On or Off. This key is available only when
Segments
key.
Mode
Notes
Dependencies /
Couplings
Preset
State Saved
Range
Key Path
1st Segment (Left or Right Limits)
1st Segment
Sets
to allow definition of the limits Left or Right of the center segment. The Left or Right
Chapter 3
Not affected by a preset but set per default to 2 MHz after a
power cycling
Saved in instrument state.
0 Hz
Frequency Span Setting
Meas Setup, Limit Masks, Limit Mask 1, Center Segment Span
Hz
Stimulus / Response
Key is only displayed when the selected segment is 1
nd
If Off the 2
limit mask segment will not be displayed even if
turned On.
If On and the Center segment is On:
st
The left hand side 1
segment is drawn to the left hand side of
the center segment at the specified power level
st
The right hand side 1
segment is drawn to the right hand side
of the center segment at the specified power
st
For both left and right 1
nd
If the 2
segment is Off the 1
limit mask span is equal to the current frequency span.
nd
If the 2
segment is On the left and right 1
specified span values.
Off
Saved in instrument state.
On | Off
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment
1st
has been selected with the
segments when On
st
segment extends so the total
st
segments use the
Mode
Return Loss
st
.
213