Agilent Technologies N1996A-503 User Reference page 31

Csa spectrum analyzer
Table of Contents

Advertisement

Key Path
Reference Level
Sets the reference level value used for amplitude measurements. The reference level is the
amplitude power or voltage represented by the top graticule line on the display. Changing
the value of the reference level changes the absolute amplitude level of the top graticule
line. The reference level (in selected amplitude units) can be set with the step keys, knob,
or numeric keypad.
Mode
Remote Command
Example
Preset
State Saved
Min
Max
Key Path
Elec Atten
Enables you to manually set the RF input attenuation level. You are able to set the input
attenuation value in 1 dB or 10 dB increments. Attenuation can be changed using the step
keys, the knob, or the numeric keypad. When using the step keys, the left and right arrows
increment the attenuation level by –1 and +1 dB respectively, and the up and down right
arrows increment the attenuation level by –10 and +10 dB respectively.
Mode
Remote Command
Remote Command Notes
Example
Chapter 3
AMPTD Y Scale, Autoscale
SA
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
:DISP:WIND:TRAC:Y:RLEV 10dBm
0 dBm
Saved in State
-150 dBm
100 dBm
AMPTD Y Scale, Ref Level
All
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
When using this command, Auto Range should be set to Off
using the Auto Range command :POW:ATT:ARAN OFF
POW:ATT 6dB
Meas Off (Spectrum Analyzer) Measurement
Mode
31