15. Position the new CI reagent gas flow module on the interior side of the back panel.
16. Use a T20 Torxhead screwdriver to attach the two screws holding the CI reagent gas flow
module to the instrument.
17. Use a 1/8 to 1/16 in. reducing ferrule to attach the 1/16 in. stainless steel tubing to the
output connector of the CI reagent gas flow module.
18. Reconnect the foreline gas tube tot he CI reagent gas flow module.
19. Attach one of the CI tubes to the transfer line if needed.
Figure 463. Attaching a CI Tube to the Transfer Line
CI Tube
Transfer Line
Tubing
Connector
20. Look through the top of the instrument and confirm that the CI tube is attached to the
mixing chamber and manifold wall.
Figure 464. Attaching a CI Tube to the Mixing Chamber and Manifold Wall
Small Glass
Manifold
Cover
CI Tube
Manifold
Wall