Disconnecting The Vacuum Interlock Cable; Removing The Vacuum Interlock Shield - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Manifold
260
TSQ 9000 Hardware Manual
3. Disconnect the vacuum interlock cable from the connector.
Figure 274. Disconnecting the Vacuum Interlock Cable
4. Pinch together the two springs on the left side the vacuum interlock shield and remove it.
Figure 275. Removing the Vacuum Interlock Shield
Springs
5. Lift the vacuum interlock shield off the instrument. The notch at the bottom of the track
will go around the handle. Be sure to straighten the ball valve handle prior to removing
the VPI shield.
Vacuum
Interlock
Cable
Connector
Ball Valve
Handle
Thermo Scientific