Thermo Scientific
Figure 338. Removing the Large Glass Manifold Covers
CAUTION - ELECTRICAL HAZARD Avoid dropping screws between chassis and side cover
or board. If you drop a screw, find it before powering the instrument on again.
12. Look into the instrument from the left side and use a T20 Torxhead screwdriver to
remove the three screws around the 4-pin feedthrough.
Figure 339. Removing the 4-pin Feedthroughs
Screws
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Large Glass
Manifold Cover
TSQ 9000 Hardware Manual
317