Turbomolecular Pump Shuts Off While In Use - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Turbomolecular pump shuts off while in use.

Table 23. Possible Remedies when Turbomolecular Pump Shuts Off While in Use
Run Leak Check diagnostics to look for leaks and address them. Excessive gas flow may cause the pump to overheat.
Ensure the temperature of the lab environment is 15-35 °C.
Make sure the flow of carrier gas to the TSQ 9000 is less than 10 mL/min.
Thermo Scientific
2
Troubleshooting
Investigating Vacuum Issues
TSQ 9000 Hardware Manual
115

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents