Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
310
TSQ 9000 Hardware Manual
14. Insert a new RF board by aligning the connectors on the bottom of the board with the
connectors on the distribution board beneath it. It will drop into the slot when it is in the
right position.
Figure 332. Reinstalling the RF Boards
15. Push the board down into the connectors.
16. Gently push the RF board toward the instrument, carefully aligning the board over the
feedthrough pins, until it stops.
17. Reattach the screw holding the RF board to the bracket.
Note This screw is required for good ground contact.
18. Reattach the top cover panel and tighten the screw holding it in place.
19. Restart your system by following the instructions in
Note The new RF board will have a slightly different performance than your current
board. It will need to be adjusted by a Field Service Engineer because the adjustments
must be made while the system is running.
Front RF
Board
Connectors
Rear RF
Board
Connectors
"Restarting the System"
on
Thermo Scientific
page
145.