3
Advanced Troubleshooting
Replacing the Power Supplies
364
TSQ 9000 Hardware Manual
7. Disconnect the large feedthrough cables labeled Dynode and Electron Multiplier.
Figure 387. Disconnecting the Electron Multiplier Cables
Electrometer
Board Cable
8. Disconnect the cables from the clips attached to the chassis.
Dynode
Electron
Multiplier
Electron
Multiplier Plate
Thermo Scientific