Thermo Scientific TSQ 8000 Preinstallation Manual
Hide thumbs Also See for TSQ 8000:

Advertisement

Quick Links

TSQ 8000
Preinstallation Requirements Guide
1R120505-0001 Revision D April 2014

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TSQ 8000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Thermo Scientific TSQ 8000

  • Page 1 TSQ 8000 Preinstallation Requirements Guide 1R120505-0001 Revision D April 2014...
  • Page 2 © 2014 Thermo Fisher Scientific Inc. All rights reserved. TSQ 8000, ISQ, TRACE 1300, TRACE 1310, TriPlus RSH, and TraceFinder are trademarks, and Xcalibur is a registered trademark of Thermo Fisher Scientific in the United States. The following are registered trademarks in the United States and other countries: Microsoft, Windows, Office, Skydrive, and Excel are registered trademarks of Microsoft.
  • Page 3 Dear Customer: Read the TSQ 8000 Preinstallation Requirements Guide, and then print and complete the following installation request form. After all items on the form are fulfilled, sign and date the form. Then, scan and email, mail, or fax this form to your local Thermo Fisher Scientific sales/service office.
  • Page 5 Offices for Thermo Scientific CMD Products For up-to-date contact information, visit www.thermoscientific.com. North America France (Also representing French-speaking North Africa, United States Algeria, Morocco, and Tunisia) 1400 North Point Pkwy #10 16 Avenue du Québec West Palm Beach, FL 33407...
  • Page 6 Offices for Thermo Scientific CMD Products—Continued Europe—continued Building 6 Sweden, Norway, and Finland No.27 Xin Jinqiao Road Pudong District Pyramidbacken 3 Shanghai 201206 S-141 75 Kungens Kurva (Stockholm) Phone......[86] (21) 6865 4588 Sweden Fax ......[86] (21) 6445 7830 E-mail: service.sid.nordic@thermofisher.com Phone......[46] (0) 8 556 468 20...
  • Page 7: Table Of Contents

    Hydrogen Safety Precautions ........xvii Using Hydrogen with TSQ 8000 ....... .xviii Hydrogen Connection Guidelines .
  • Page 8 TraceFinder 3.2..........25 viii TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 9 IEC 61010-2-081:2001 • ISQ standards: EMC EN 61326-1:2006. Safety IEC 61010-1:2001, IEC 61010-2-081:2001 • TSQ 8000 standards: EMC EN 61326-1:2006. Safety IEC 61010-1:2001, IEC 61010-2-081:2001 Low Voltage Safety Compliance This device complies with Low Voltage Directive 2006/95/EC and harmonized standard EN 61010-1:2001.
  • Page 10 Notice on Lifting and Handling of Thermo Scientific Instruments For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher Scientific instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/or bulky for one person alone to handle safely.
  • Page 11 For manufacturing location, see the label on the instrument. WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union (EU) Member State, and these companies should dispose of or recycle this product.
  • Page 12 Conformité DEEE Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant: Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union européenne et ce produit devrait être collecté...
  • Page 13: Preface

    • Contacting Us About Your System Thermo Scientific systems provide the highest caliber gas chromatography/mass spectrometry (GC/MS) instrumentation available on today’s market. GC/MS represents a combination of two powerful analytical techniques: GC, which acts as a separation technique, and MS, which acts as a detection technique. Complex mixtures of individual compounds can be injected into the GC, either manually or by an autosampler and then separated for presentation to the MS.
  • Page 14: Related Documentation

    Related Documentation The TSQ 8000 system includes Help and these manuals as PDF files: • TSQ 8000 Preinstallation Guide, PN 1R120505-0001 • TSQ 8000 User Guide, PN 1R120505-0002 •...
  • Page 15: System Requirements

    • Thermo Scientific™ Xcalibur™ Foundation • Thermo Scientific™ TraceFinder™ Check release notes for compatibility with TSQ 8000 instrument control software. Check release notes for compatibility with Thermo Foundation, Xcalibur, and TSQ 8000 instrument control software. Safety and Special Notices Make sure you follow the precautionary statements presented in this guide. The safety and other special notices appear in boxes.
  • Page 16 LIFTING HAZARD: Indicates that a physical injury could or might occur if two or more people do not lift an object. MATERIAL AND EYE HAZARD: Indicates that eye damage could or might occur. RADIOACTIVE HAZARD: Indicates that exposure to radioactive material could or might occur. TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 17: Hydrogen Safety Precautions

    You should also consider purging the hydrogen to further reduce hazards and ensure anyone who will be working with hydrogen has basic hydrogen safety training. Thermo Scientific TSQ 8000 Preinstallation Requirements Guide xvii...
  • Page 18: Using Hydrogen With Tsq 8000

    Using Hydrogen with TSQ 8000 To use hydrogen with the TSQ 8000, you must always shut off the GC carrier gas before venting or turning off the TSQ 8000. There are three hydrogen safety screws on the TSQ 8000 that must be in place.
  • Page 19: Hydrogen Connection Guidelines

    Preface Hydrogen Safety Precautions • All the covers and panels of the TSQ 8000 system are firmly attached. • The vent valve is tightly closed if you vented the system. • All fittings, ferrules, and o-rings are sealed. Hydrogen Connection Guidelines Use the following guidelines to safely connect hydrogen to your system: •...
  • Page 20: Purchasing Hydrogen

    There is a loss of flow to the ventilation system – A hydrogen detector alarms at 25% of the lower flammable limit of hydrogen in air. The oxygen exhausted by the electrolyzer should be vented to the outside as well. TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 21: Properly Storing Hydrogen

    Bulk hydrogen systems represent a level of complexity again which is beyond the scope of this document; however some general guidelines are provided. Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 22 • As liquid hydrogen systems also have a cryogenic hazard, additional safety considerations for the use of cryogenic liquids might be necessary. xxii TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 23: Hydrogen Safety Codes, Standards And References

    • NFPA 55, Standard for the Storage, Use, and Handling of Compressed Gases and Cryogenic Fluids in Portable and Stationary Containers, Cylinders, and Tanks, 2005 Edition • NFPA 68, Standard on Explosion Protection by Deflagration Venting • NFPA 70, National Electrical Code Thermo Scientific TSQ 8000 Preinstallation Requirements Guide xxiii...
  • Page 24: Hazardous Substances Precautions

    When working with biohazardous materials, you are responsible for fulfilling the following mandatory requirements: • Providing instructions on how to safely handle biohazardous material. • Training operators to be aware of potential hazards. xxiv TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 25: Venting Toxic Gases

    Go to www.thermoscientific.com/wps/portal/ts/contactus.  To suggest changes to documentation or to Help • Fill out a reader survey online at www.surveymonkey.com/s/PQM6P62. • Send an e-mail message to the Technical Publications Editor at techpubs-austin@thermofisher.com. Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 27: Site Preparation

    Site Preparation This chapter describes how to prepare your site before the Thermo Scientific Field Service Engineer arrives to install the TSQ 8000. Contents • Entrance Requirements • Workbench and Space Requirements • Lighting Requirements • Power Requirements • Environmental Requirements •...
  • Page 28: Entrance Requirements

    TriPlus 300 Headspace Autosampler (Box 1 of 2) TriPlus 300 Headspace Autosampler (Box 2 of 2) The computer, keyboard, monitor, foreline pump, and TSQ 8000 Installation Kit are included in the box with the TSQ 8000. TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 29: Workbench And Space Requirements

    Site Preparation Workbench and Space Requirements Workbench and Space Requirements Use the following guidelines to ensure you have enough space to set up the TSQ 8000 system. Figure 1 Figure 2 for system orientation. 1. Ensure you have adequate workbench space for the system. Refer to the table below for exact measurements of each component.
  • Page 30: Tsq 8000 89 35 40 16 45

    3. Make sure you have at least 91 cm (3 ft.) of clearance above the system. This space allows room for optional accessories (such as autosamplers) and proper heat dissipation. 4. Make sure your workbench can support a standard TSQ 8000 system. Keep in mind, additional instruments add to the total weight.
  • Page 31: Lighting Requirements

    120 Vac +6/-10% or 230 Vac ±10%. Note Due to the power draw of the TSQ 8000 and GC, each instrument must be on its own dedicated circuit. These two circuits must be separate from the circuits used for the computer and other equipment such as the autosampler.
  • Page 32: Triplus 300 Headspace

    This instrument must be on its own circuit. Foreline pump and optional mech pump plug into the TSQ 8000 MS. At startup, the foreline pump and optional mech pump draw an additional 30.8 A at 120 Vac and an additional 17 A at 230 Vac.
  • Page 33 • Included power cords are 2 m (6 ft) long . Contact your local Customer Service office to discuss power cordset concerns. 3. The TSQ 8000 comes with the required number and type of power cords for your region. Table 4 will help you identify the correct power cord for your region and instrument.
  • Page 34 C13 Power Cords (For PC, Monitors, C19 Power Cords (For GC Region and Autosamplers) and MS) China 250V Europe-Schuko 250V United Kingdom 250V Danish 250V Italian 250V Israeli 250V Indian 250V Argentina 250V TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 35: Environmental Requirements

    IMPORTANT If you are not operating the TSQ 8000, it is safe to leave the instrument powered on in a room with a temperature between 5 and 40 °C.
  • Page 36 2. Ensure that the relative humidity in your laboratory is between 5 and 95% with no condensation. A temperature and humidity monitor in your laboratory helps ensure that the climate is within these specifications. TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 37: Gas Equipment Requirements

    Note Thermo Fisher Scientific installation specifications require helium as a carrier gas and argon as a collision gas. You must have one tank of each gas at installation, or the field service engineer will not be able to run specifications on your TSQ 8000. Table 6. Carrier Gas Specifications...
  • Page 38: Ci Gas

    Turn off the supply until the GC column is in the inlet of the TSQ 8000. Whenever you use hydrogen, it is critical to test all connections, lines, and valves for leaks before using the instrument. When performing maintenance, be sure to turn off the hydrogen supply.
  • Page 39: Collision Gas

    CI reagent gas flow module. Collision Gas • You may use argon or nitrogen as collision gases for the TSQ 8000. Typical cylinders are about 23 cm (9 in.) in diameter by 140 cm (55 in.) tall and output >15,000 kPa (>2200 psig).
  • Page 40: Other Gas Specifications

    Requirements Guide for more information. Other Gas Specifications • If your TSQ 8000 system will be equipped with a Direct Insertion Probe, make sure you have compressed air, which is used to cool the probe. • If you have a TriPlus autosampler with the SPME conditioning station or a TriPlus Headspace autosampler, you need to obtain a low-pressure, single-stage regulator (0-30 psi) for nitrogen purging.
  • Page 41: Using Hydrogen

    Receiving Requirements Using Hydrogen To safely use hydrogen in your TSQ 8000, you should have a hydrogen sensor installed in your GC. Field Service Engineers can install a sensor. but they are not authorized to install or repair any instrument using hydrogen as a carrier gas unless the instrument is equipped with the appropriate sensor.
  • Page 42: What Happens Next

    “End User License Agreements” page 17 and fill out the TSQ 8000 Installation Request Form. After the Installation Request Form is received, Customer Support will contact you to schedule the installation of your system. It is important to confirm that all the requirements on the form are met BEFORE the Field Service Engineer arrives.
  • Page 43 OpenOffice what it is today - the world's leading open-source office software. Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 44 Legal Entity authorized to submit on behalf of the copyright owner. For the purposes of this definition, "submitted" means any form of electronic, verbal, or written communication sent to the Licensor or its representatives, including but not limited to TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 45 Derivative Works; or, within a display generated by the Derivative Works, if and wherever such third-party notices normally appear. The contents of the NOTICE file are for informational purposes only and do not Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 46 Works thereof, You may choose to offer, and charge a fee for, acceptance of support, warranty, indemnity, or other liability obligations and/or rights consistent with this License. However, in accepting such obligations, You may act only on Your own behalf and on Your sole TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 47 4. No title, ownership or interest in the Software or Documentation is transferred to Customer by this Agreement or by the payment of any fee. 5. Any rights not expressly granted herein to Customer are reserved to Thermo and Information Partners. C. WHAT CUSTOMER MAY DO Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 48 Site for the reconstruction of any lost data. It is solely Customer’s responsibility to evaluate the TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 49 Thermo will be for acknowledgment purposes only and none of the terms set forth in the purchase order will be binding upon Thermo. Any representation, promise, warranty, covenant or Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 50 (under FAR 52.227019 and DFARS 252.227-7013 or parallel regulations). The manufacturer for this purpose is Thermo Finnigan LLC, 355 River Oaks Parkway, San Jose, CA 95134-1908 or the applicable Information Partner. TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 51 END-USER LICENSE AGREEMENT (EULA) FOR THERMO FISHER SCIENTIFIC TRACEFINDER 3.2 AND MICROSOFT OFFICE PROFESSSIONAL PLUS 2013 Thermo Scientific TraceFinder 3.2 with Microsoft Office Professional Plus 2013 Edition Licenses: One (1) for Microsoft Office Professional Plus 2013 Edition and two (2) for Thermo Fisher Scientific TraceFinder 3.2.
  • Page 52 What does this agreement apply to? This agreement applies to the software, the media on which you received the software, and also any Microsoft updates, supplements, and services for the software, unless other terms come with them. TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 53 As an exception, you may allow others to access the software simultaneously only to provide you with technical support. B. BINDING ARBITRATION AND CLASS ACTION WAIVER Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 54 $75,000 or less whether or not you are an individual or how you use the software, the AAA Supplementary Procedures for Consumer-Related Disputes will also apply. To TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 55 Section B remaining in full force and effect. 10. Third-Party Beneficiary. Microsoft Corporation is not a party to this agreement but is a third-party beneficiary of your and the licensor’s agreement to resolve disputes through Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 56 4. Updates and Upgrades. You may only obtain updates or upgrades for the software from Microsoft or authorized sources. Certain upgrades, support, and other services may be offered only to users of genuine Microsoft software. To identify genuine Microsoft software, see howtotell.com. E. INTERNET-BASED FEATURES; PRIVACY TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 57 They also can be used to digitally sign files and macros to verify the integrity and origin of the file contents. The software retrieves certificates and updates certificate revocation lists using the Internet, when available. Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 58 (iv) create obscene or scandalous works using the media elements. Other media elements, which are accessible on Office.com or on other websites through features of the software, are governed by the terms on those websites. TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 59 FOR ADMINISTRATIVE PURPOSES, TO STORE CONFIGURATION DATA, ENGINEERING AND/OR CONFIGURATION TOOLS, OR OTHER NON-CONTROL APPLICATIONS, THE FAILURE OF WHICH WOULD NOT RESULT IN DEATH, PERSONAL INJURY, OR SEVERE PHYSICAL OR ENVIRONMENTAL DAMAGE. THESE NON-CONTROLLING APPLICATIONS Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 60 Microsoft, Outlook, SkyDrive, and Windows are registered trademarks of Microsoft Corporation in the United States and/or other countries. END-USER LICENSE AGREEMENT – THERMO FISHER SCIENTIFIC TRACEFINDER 3.2 TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 61 1. Create derivative works of, decompile, reverse engineer or attempt to determine the source code of the Software. 2. Sell, distribute or commercially exploit the Software and Documentation without the prior written consent of Thermo. Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 62 DETERMINED WITHOUT REGARD TO WHEN THE CUSTOMER SHALL HAVE LEARNED OF THE ALLEGED DEFECT, INJURY, OR LOSS. 5. IN NO EVENT SHALL THE TOTAL LIABILITY OF INFORMATION PARTNERS AND THERMO EXCEED THE LICENSE FEE PAID BY CUSTOMER FOR THE TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...
  • Page 63 Agreement will remain in full force and effect. J. AUDIT 1. Customer grants Thermo the right to audit, during regular business hours, Customer’s use of the Software and Documentation to ensure compliance with this Agreement. Thermo Scientific TSQ 8000 Preinstallation Requirements Guide...
  • Page 64 U.S. Government is subject to the restricted rights applicable to commercial computer software (under FAR 52.227019 and DFARS 252.227-7013 or parallel regulations). The manufacturer for this purpose is Thermo Finnigan LLC, 355 River Oaks Parkway, San Jose, CA 95134-1908 or the applicable Information Partner. TSQ 8000 Preinstallation Requirements Guide Thermo Scientific...

Table of Contents