5. Twist and remove the source retainer magnets that hold the magnet yoke to the back of
the manifold door. The magnet yoke will pull free of the door, releasing the alignment
collar.
Figure 446. Removing the Source Retainer Magnets
Magnet Yoke
Source Retainer
Magnets
6. Set the old parts on a clean surface temporarily. There is a place to store them in the
Advanced EI source kit.
7. Remove the new source magnets and magnet yoke from the Advanced EI source kit and
put them together as shown in
Figure 447. Assembling the Advanced EI Source Magnets and Yoke
Magnet Yoke
8. Attach the source magnets and yoke to the manifold door using the new tray alignment
pins in the Advanced EI source kit. See
Figure
447.
Figure
448.
4
Upgrade Equipment
Advanced EI Ion Source
Advanced EI
Source
Magnets
TSQ 9000 Hardware Manual
421