Replacing The Back Manifold Plate - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
11. Remove the tray alignment pin and manifold o-ring and put them on the new back
manifold plate. Replace the pin and o-ring, if necessary.
Figure 293. Replacing the Back Manifold Plate
Manifold
O-Ring
Alignment
Pin
12. Reattach the back manifold plate to the instrument. The back manifold plate must be
reinstalled with the pin at the bottom.
13. Slide the power supply back into the instrument.
Note Make sure the electrometer board cable is not in the way when you reinstall the
power supply.
14. Reattach the two screws holding the power supply to the instrument.
15. Reconnect the power cables on the left hand sub panel of the TSQ 9000 instrument.
16. Replace the analyzer tray into the instrument by following the instructions in
the Analyzer Tray in the TSQ 9000 Instrument"
17. Reattach the right side panel and top cover panel and tighten the screws.
18. Reattach the GC oven and transfer line.
19. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing Components of the Manifold
Back
Manifold
Plate
on
page
142.
"Restarting the System"
TSQ 9000 Hardware Manual
"Replacing
on
page
145.
271