Figure 133. Removing the Collision Cell and Quadrupole 3 from the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
9. Replace the collision cell assembly if needed.
Note If you do not need to replace quadrupole 1, go to
Figure 134. Removing Screws from the Quadrupole Rod Endcaps
Collision
Cell
Assembly
10. Slide the Q1 quad assembly away from the Q1 entrance lens and endcap.
11. Replace Q1 if necessary.
12. Slide Q1 back into the Q1 entrance lens endcap.
13. Slide the collision cell assembly onto the analyzer tray. Hold the collision cell and Q1 in
the tray and make sure Q1 and the collision cell are correctly positioned before tightening
the screws on the Q1 exit endcap.
Replacing Components of the Analyzer
step
14.
Screw to
Q1 Entrance Lens
Grounding
Endcap Screws to
Wire
Analyzer Tray
TSQ 9000 Hardware Manual
3
Advanced Troubleshooting
Q3 Exit Lens
Assembly
Screws
Q1 Quad
Assembly
Q1 Entrance
Lens Endcap
161