22. Pull the tubing out of the manifold. Replace if necessary.
23. Disconnect the collision gas module cable from the distribution board.
Figure 414. Disconnecting the Collision Gas Module Cable from the Distribution Board
24. Disconnect the 1/16 in. tubing from the bulkhead compression fitting located on the left
hand sub panel.
Figure 415. Disconnecting the Cable to the Bulkhead Fitting
Bulkhead
Fitting
25. Use a T20 Torxhead screwdriver to remove the two screws on the collision gas module.
26. Remove the collision gas module and replace if necessary.