Repeller Plate - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Ion Source

Repeller Plate

186
TSQ 9000 Hardware Manual
To replace the repeller plate:
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
4. Open the manifold door.
Figure 168. Opening the Manifold Door
"Shutting Down the TSQ 9000 System"
Screws
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents