Lens Driver Board

Thermo Scientific

To replace the lens driver board

Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127. You do not have to remove the ion source cartridge or the column.
2. Slide the top cover panel toward the back of the instrument and it them off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 340. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
4. Lift the left hand front panel off the instrument.
3
Replacing the Boards in the TSQ 9000 Instrument
"Shutting Down the TSQ 9000 System"
Screws
TSQ 9000 Hardware Manual
Advanced Troubleshooting
Screw on Front
of Instrument