27. Reconnect the collision gas module to the instrument with the two screws.
28. Reconnect the 1/16 in. tubing to the bulkhead fitting through the a graphite Vespel
ferrule.
29. Reconnect the collision gas tubing to the collision gas module at the compression fitting.
30. If you removed it from the manifold, replace the collision gas tubing through the
compression fitting in the manifold.
31. Reconnect the collision gas tubing to the collision cell.
32. Replace the large glass manifold cover.
33. Reconnect the collision gas module cable to the distribution board.
34. Reconnect the tubing to the tee connected to the foreline adapter.
35. Replace the controller interface board and PC communication board and reconnect the
Ethernet cable.
36. Replace the rear RF board.
37. Replace the top, left hand front, and right side panels and secure with screws.
38. Move the TSQ 9000 next to the GC.
39. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing the Collision Gas Components
"Restarting the System"
on
TSQ 9000 Hardware Manual
page
145.
Electrical
Cable
389